久久久久久久网-久久久久久久男人的天堂-久久久久久久免费视频-久久久久久久久综合影视网-尤物视频网站在线观看-尤物视频网站在线

【Tech Online】HCP ELISA Method Validation – Dilution Linearity


Introduction

Host Cell Protein (HCP) ELISA detection methods must undergo method validation in accordance with regulatory-approved experimental protocols (such as ICH or FDA guidelines) to ensure they produce accurate results. Essentially, obtaining accurate HCP ELISA results requires two key conditions: the HCP concentration in the sample must be above the kit’s Limit of Quantitation (LOQ), and the detection antibodies for HCP must be in excess. To meet these two conditions, dilution linearity validation of test samples (process samples and final drug substance bulk) is imperative. Therefore, dilution linearity is one of the critical experiments for verifying the specificity and accuracy of HCP ELISA methods.

 

There are two main reasons for poor dilution linearity in HCP detection:

1. The excess antibody condition is not satisfied for certain HCPs in the sample, requiring further dilution to reach the "Minimum Required Dilution" (MRD);

2. The product protein itself or certain components in the product formulation buffer interfere with HCP detection. In such cases, samples can be treated using designated sample diluents (see end of article) or buffer exchange to meet detection requirements.

 

Regarding sample diluents, the optimal choice is to use the same diluent as that for the standard. Cygnus provides corresponding sample diluents for each ELISA detection method to ensure the accuracy and reproducibility of test results. For users of Cygnus HCP ELISA kits, we recommend defining acceptable dilution linearity as: "When performing two-fold dilutions, the corrected HCP concentration does not vary by more than ±20%". Additionally, detection values should be avoided from being close to the LOQ—specifically, the HCP detection value of the diluted sample must not be lower than twice the LOQ.

 

In the following sections, this article will provide a detailed explanation of the operating procedures and precautions for dilution linearity in HCP ELISA method validation.

 

Why Establish Dilution Linearity?

When testing a new HCP ELISA method, establishing dilution linearity is the first fundamental experiment. Its purpose is to confirm whether the selected HCP ELISA is suitable for detecting HCPs in your process samples and drug substance bulk. Once dilution linearity is confirmed, it provides the experimental condition information required for further analytical validation—hence, dilution linearity is typically the first validation experiment to be conducted.

 

Specifically, the objective of the dilution linearity experiment is to determine the reaction curve and full quantitation range of the ELISA method for HCPs in the sample. Successful validation of dilution linearity indicates that:

· The excess antibody condition is satisfied for all types of HCPs present in the sample;

· The sample matrix does not interfere with HCP quantitation.

 

Together, these ensure the reliability of HCP ELISA detection results.

 

How to Perform a Dilution Linearity Experiment?

Prior to ELISA analysis, any sample that may exceed the Upper Limit of Quantitation (ULOQ) should first undergo a dilution linearity experiment. The process involves the following steps:

1. Perform multiple gradient dilutions of the HCP test sample using a validated sample diluent;

2. Detect these diluted samples;

3. Calculate the corrected HCP concentration for each diluted sample (i.e., measured HCP concentration × dilution factor).

 

In most cases, after reaching a certain dilution factor, the corrected HCP concentration of the diluted sample will stabilize at a relatively constant value. This dilution factor is referred to as the "Minimum Required Dilution" (MRD).


Figure 1. Minimum Required Dilution (MRD)



As shown in the example in the table below, the sample did not exhibit good dilution linearity at high concentrations (undiluted, 1:2 dilution, and 1:4 dilution). However, with further dilution, we can determine a Minimum Required Dilution (MRD)—the dilution factor at which acceptable dilution linearity is achieved (see the blue row in the table below). By definition, acceptable dilution linearity is achieved when the change in corrected HCP concentration between two-fold dilutions does not exceed ±20%. Meanwhile, detection values should be kept away from the Limit of Quantitation (LOQ); specifically, the HCP detection value of the diluted sample must not be lower than twice the LOQ (in this example, the HCP detection value of the 1:64 diluted sample was < 2×LOQ, so it was excluded from the statistics).

 

The percentage change is calculated by subtracting the corrected concentration of the previous diluted sample from that of the current one, then dividing the difference by the corrected concentration of the previous diluted sample. For instance, the percentage change of the 1:2 diluted sample relative to the undiluted sample is calculated as follows: [(233 - 146)/146] × 100% = 60%.

Sample Dilution Gradient

HCP Dilution-Corrected Value (ng/mL)

Percentage Change from the Previous Diluted Sample

Undiluted

146

NA

1:2

233

60%

1:4

312

34%

1:8

361

16%

1:16

356

1%

1:32

370

4%

1:64

未計算

(<2LOQ)

NA

From the data in the table, we can conclude that the Minimum Required Dilution (MRD) for this sample is 1:4. Therefore, the final calculated HCP concentration should be the average of all dilution-corrected HCP concentrations that fall at or below the MRD and above the Limit of Quantitation (LOQ). (In this example, the average of 312, 361, 356, and 370—i.e., 350 ng/mL—should be used.)

 

In practical applications, we should evaluate the dilution linearity for all sample types requiring HCP detection (including process samples and the final drug substance bulk). Additionally, the MRD values for each sample type and matrix, as well as specific guidelines for diluting different samples, should be incorporated into Standard Operating Procedure (SOP) documents.

Causes of Poor Dilution Linearity and Corresponding Solutions

In HCP ELISA detection, poor dilution linearity may occur for certain HCPs, which means the "excess antibody condition" is not satisfied for those specific HCPs in the sample. For instance, individual HCPs at high concentrations may saturate the antibodies targeting them, leading to the "high-dose hook effect"—a phenomenon that results in inaccurate quantitation of the affected HCPs.


Figure 2. Schematic Diagram of the High-Dose Hook Effect

Additionally, certain hitchhiker proteins may interact with product proteins and become significantly enriched during the purification process, thereby completely interfering with the linearity of the entire assay.

Figure 3. Hitchhiker Proteins Interfering with Assay Linearity via Interaction with

Other factors that may also lead to poor dilution linearity include interference with the detection process by certain components in the product formulation (or the buffer solution of process samples). For Cygnus HCP ELISA kits, known interfering factors include extreme pH, detergents, organic solvents, high protein concentrations, and high salt concentrations. Therefore, we recommend that samples must be diluted using the same diluent as that for the standards, as detailed in the table below.

Item No.

Product Name

Application

Specification

I028

Sample Diluent

Suitable for Most HCP Kits, such as CHOF550-1,E.coliF1020, F410etc.

100 ml, 500 ml, 1000 ml

I700

Sample Diluent

Suitable for HEK 293 HCP (F650S) Kit

100 ml, 500 ml, 1000 ml

I094

Sample Diluent

Suitable for MDCK (F800) and HeLa (F810) HCP Kits

100 ml, 500 ml, 1000 ml

I600

Sample Diluent

Suitable for Most Kits in the Mix-N-Go Series, Including Protein A Detection Kits (F600, F610, F740, F910, F950, etc.) and Restriction Endonuclease Residue Detection Kit (F960)

25 ml, 100 ml

F031A

Sample Diluent

Suitable for BSA (F030) Detection Kit

100 ml, 500 ml, 1000 ml

I800

Sample Diluent

Suitable for PROchievA? Protein A (F965) Kit

25 ml, 100 ml

F223A

Sample Diluent

Suitable for NS/0 HCP (F220) Kit

100 ml, 500 ml, 1000 ml

F233A

Sample Diluent

Suitable for A549 HCP (F230) Kit

100 ml, 500 ml

F213A

Sample Diluent

Suitable for Goat IgG (F210) Detection Kit

100 ml, 500 ml, 1000 ml

F243A

Sample Diluent

Suitable for Goat Milk Total Protein (F230) Detection Kit

100 ml, 500 ml, 1000 ml

Overview of Sample Diluents for Cygnus Bioprocess Impurity Detection Kits (by Kit Type)

Finally, we emphasize again that the dose-response curve will exhibit a positive slope and the detection results will be accurate only when antibodies are in excess.

 

If poor dilution linearity is caused by insufficient HCP antibodies or sample matrix interference, the measured HCP concentration will show an upward trend as the sample dilution factor increases. Therefore, it is necessary to further dilute the sample to achieve the Minimum Required Dilution (MRD). For other causes, the accuracy of detection can also be improved by optimizing experimental procedures. If increasing the sample dilution factor still fails to resolve the issue of poor dilution linearity, there may be a problem with the purification process.

 

Cygnus provides a corresponding Qualification Summary (QS) document for each type of detection kit. If needed, please click here to fill in the information to request it, or contact Ximeijie (the exclusive distributor of Cygnus in China). If you encounter dilution linearity-related issues when using Cygnus HCP ELISA kits and cannot resolve them through the aforementioned methods, you are also welcome to contact Ximeijie at any time—our technical support team will provide corresponding solutions and suggestions for your specific problems. In the future, we will continue to share a series of content on HCP ELISA method validation, including spike recovery and HCP antibody suitability validation.

 

 

For over 25 years, Cygnus Technologies has specialized in providing products and analytical methods for the biotechnology and biopharmaceutical industries, with the aim of accelerating the research and development (R&D) phase and enhancing product quality. Cygnus develops and manufactures bioprocess residue detection kits, which are used to detect specific impurities across more than 50 different expression systems. As an expert in high-sensitivity analytical technologies focused on immunoassays for biotechnological applications, Cygnus’ products and services are utilized by over 95% of biopharmaceutical companies worldwide and have gained broad recognition from regulatory authorities such as the FDA (U.S. Food and Drug Administration), NMPA (National Medical Products)

As the exclusive distributor of Cygnus in China, Beijing XMJ Technology Co., Ltd. has established long-term and stable cooperative relationships with numerous well-known domestic pharmaceutical companies and CRO/CMO enterprises. Over the years, Ximeijie’s products and services have helped many enterprises accelerate the R&D phase, enhance drug quality, purity and safety, optimize R&D processes more efficiently, shorten time-to-market for products, and reduce QC costs. If you are interested in the aforementioned products, please feel free to send email to info@xmjsci.com or visit the website www.gp82.cn for more information.


欧美激情一区二区三区在线 | 99热精品在线| 九九久久国产精品| 日韩中文字幕一区| 香蕉视频久久| 国产不卡福利| 99久久精品费精品国产一区二区| 免费国产在线视频| 亚洲第一色在线| 久久精品成人一区二区三区| 色综合久久手机在线| 欧美国产日韩在线| 四虎影视库国产精品一区| 毛片电影网| 欧美激情影院| 亚洲天堂免费| 成人a大片在线观看| 国产亚洲精品aaa大片| 999久久久免费精品国产牛牛| 四虎久久精品国产| 青青久久国产成人免费网站| 一级毛片看真人在线视频| 好男人天堂网 久久精品国产这里是免费 国产精品成人一区二区 男人天堂网2021 男人的天堂在线观看 丁香六月综合激情 | 999精品影视在线观看| 99色视频在线观看| 99热精品在线| 一级女人毛片人一女人| 欧美18性精品| 成人免费网站久久久| 国产不卡在线观看| 午夜激情视频在线播放| 日韩专区第一页| 亚欧乱色一区二区三区| 999久久狠狠免费精品| 精品久久久久久综合网| 麻豆污视频| 99久久精品国产麻豆| 国产亚洲免费观看| 精品国产一区二区三区久久久狼| 免费一级生活片| 天天色色色| 精品久久久久久免费影院| 日本特黄特色aa大片免费| 欧美日本二区| 天天做日日爱| 好男人天堂网 久久精品国产这里是免费 国产精品成人一区二区 男人天堂网2021 男人的天堂在线观看 丁香六月综合激情 | 九九久久99综合一区二区| 99色视频在线观看| 国产成人精品综合| 91麻豆精品国产高清在线| 欧美a免费| 欧美另类videosbestsex高清| 欧美另类videosbestsex视频| 999久久狠狠免费精品| 午夜在线亚洲| 日韩专区亚洲综合久久| 亚欧乱色一区二区三区| 精品国产一区二区三区国产馆| 欧美1区2区3区| 国产视频一区二区在线观看 | 欧美日本免费| 韩国毛片| 午夜欧美成人久久久久久| 成人免费观看的视频黄页| 日韩av东京社区男人的天堂| 国产一区二区福利久久| 99热精品一区| 亚欧成人毛片一区二区三区四区 | 日韩在线观看网站| 欧美日本免费| 中文字幕97| 日本在线不卡视频| 成人免费观看视频| 日韩男人天堂| 久久精品大片| 韩国三级一区| 色综合久久天天综线观看| 99久久精品国产高清一区二区| 好男人天堂网 久久精品国产这里是免费 国产精品成人一区二区 男人天堂网2021 男人的天堂在线观看 丁香六月综合激情 | 一级毛片看真人在线视频| 夜夜操天天爽| 二级特黄绝大片免费视频大片| 成人影视在线观看| 国产麻豆精品免费视频| 精品久久久久久免费影院| 韩国毛片| 欧美大片a一级毛片视频| 高清一级做a爱过程不卡视频| 99久久精品国产麻豆| 精品国产亚洲一区二区三区| 欧美a级大片| 国产视频一区二区在线播放| 国产不卡高清在线观看视频| 毛片高清| 日韩免费在线| 日本免费乱人伦在线观看| 四虎影视库| 香蕉视频一级| 麻豆系列 在线视频| 青青青草影院 | 久久国产一区二区| 日韩一级黄色片| 欧美国产日韩久久久| 一级女人毛片人一女人| 国产视频久久久| 国产麻豆精品视频| 毛片成人永久免费视频| 国产视频一区在线| 91麻豆精品国产高清在线| 精品国产香蕉在线播出| 日韩男人天堂| 久久精品免视看国产成人2021| 欧美a级大片| 在线观看成人网| 91麻豆tv| 欧美日本二区| 国产高清视频免费| 精品视频在线观看一区二区| 深夜做爰性大片中文| 欧美爱色| 九九精品在线| 国产成a人片在线观看视频| 欧美电影免费看大全| 日本久久久久久久 97久久精品一区二区三区 狠狠色噜噜狠狠狠狠97 日日干综合 五月天婷婷在线观看高清 九色福利视频 | 欧美a级片免费看| 成人av在线播放| 美国一区二区三区| a级精品九九九大片免费看| 国产麻豆精品免费密入口| 午夜精品国产自在现线拍| 精品国产三级a| 成人影院一区二区三区| 成人a大片高清在线观看| 韩国毛片基地| 免费毛片基地| 亚欧乱色一区二区三区| 久久国产精品只做精品| 国产一区二区精品| 欧美激情一区二区三区视频 | 色综合久久天天综线观看| 午夜在线影院| 韩国三级视频在线观看| 九九干| 日韩专区第一页| 精品国产一区二区三区久久久蜜臀| 国产不卡高清在线观看视频| 欧美国产日韩精品| 国产成人啪精品| 四虎影视久久久| 韩国三级视频网站| 色综合久久天天综合| 九九九国产| 黄视频网站在线看| 欧美另类videosbestsex高清 | 日韩欧美一及在线播放| 成人影视在线观看| 一级毛片看真人在线视频| 久久国产一久久高清| 亚欧乱色一区二区三区| 韩国三级一区| 免费毛片播放| 国产网站在线| 麻豆污视频| 国产91丝袜高跟系列| 日日爽天天| 精品国产亚洲人成在线| 国产91丝袜在线播放0| 日韩欧美一二三区| 免费的黄视频| 午夜精品国产自在现线拍| 91麻豆精品国产自产在线| 欧美激情一区二区三区在线播放 | 四虎影视库| 久久精品欧美一区二区| 免费毛片播放| 久久99欧美| 天天做人人爱夜夜爽2020毛片| a级黄色毛片免费播放视频| 香蕉视频一级| 亚洲女人国产香蕉久久精品| 精品久久久久久综合网| 国产一级强片在线观看| 国产不卡在线观看视频| 久久国产一久久高清| 国产不卡福利| 99久久精品国产高清一区二区| 99色视频| 免费国产一级特黄aa大片在线| 青青久久网| 一级女人毛片人一女人| 国产a视频精品免费观看| 欧美国产日韩一区二区三区| 黄色短视频网站| 日本特黄特黄aaaaa大片| 国产不卡在线观看| 九九久久99| 国产国语在线播放视频| 精品视频免费看| 国产91精品系列在线观看| 尤物视频网站在线观看|